Protein Sample Preparation and Analysis Application Manual - Technical Appendices
  Protein Sample Preparation and Analysis Application Manual - Technical Appendices


Evaluation of Chromatography Column Packing Efficiency
Select the link below to access this Appendix item.

Top


Optional Pre-Treatment to Improve Recovery of Samples from Ultrafiltration Spin Filters
Poor sample recovery may be experienced with the Omega™ UF membranes and devices because of non-specific adsorption. It is possible to pre-treat these membranes and devices with protein or detergents to reduce the potential for non-specific adsorption of the target analyte. A protocol is outlined below.

Protocol for Optional Pre-Treatment to Improve Recovery of Samples from Ultrafiltration Spin Filters

  1. Add sufficient volume of one of the following:
    1. 1% (w/v) BSA
    2. 1% (w/v) IgG in phosphate buffered saline (PBS)
    3. 5% (w/v) SDS
    4. 5% (v/v) Tween-20
    5. 5% (v/v) Triton*-X100
    6. 5% (w/v) PEG compound in high purity water to partially fill the UF device.
  2. Close the cap and mix by inversion to coat the solution over the internal surfaces of the device.
  3. Incubate the solution in the device for at least one hour at room temperature.
  4. Discard the above solution by pouring it out of the device, and rinse the device thoroughly with high purity water.
    Tip: Rinsing to remove detergents such as SDS, Tween, and Triton may require more than the above wash steps.
  5. To ensure that pre-treatment solution is removed from the membrane, add high purity water to the device and spin at the recommended speed for each device to pass the fluid through the UF membrane. Discard the filtrate.
  6. The device can be used immediately or stored for later use. If the device is to be used later, add sterile high purity water (rendered sterile by autoclaving or filtration with a 0.2 µm sterilizing grade filter [see Acrodisc® syringe filters with Supor® PES membrane for sterile filtration (PN 4612)] to the sample reservoir and store at 4 °C to retard bacterial growth.
    Tip: Do not allow the membrane to dry out once the device has been pre-treated.

Top


Using Pall's Vacuum Manifold with Multi-Well Filter Plates
Select the link below to access this Appendix item.

Top


General Filtration Procedures for AcroPrep™ and AcroWell™ Filter Plates
Select the link below to access this Appendix item.

Top


Chromatography Products Selection Guide
Select the link below to access this Appendix item.

Top


Base Materials for BioSepra Media Chromatography Resins

Top


Characteristics of Pall's Protein Sample Preparation and Analysis Media
Select the link below to access this Appendix item.

Top


Chemical Compatibility Guide for Pall's Protein Sample Preparation and Analysis Media

Top