Materials of Construction
Filter Media: Omega membrane (low protein-binding, modified polyethersulfone)
Sample Reservoir and Filtrate Receiver: Polysulfone
Sample Reservoir Cap: Polyethylene
Insert Without Membrane: High density polyethylene
Filtrate Receiver Cap and Insert Release: Polypropylene
Effective Filtration Area
15.2 cm2
Dimensions
Outside Diameter (maximum): 6 cm (2.4 in.)
Overall Height (fully assembled with cap): 11.3 cm (4.5 in.)
Capacities
Maximum Sample Volume: 60 mL
Final Concentrate Volume: 3.5 - 4 mL
Maximum Filtrate Receiver Volume: 60 mL
Hold-up Volume (membrane/support): 0.2 mL
Operating Temperature Range
0 - 40 °C (32 - 104 °F)
pH Range
1 - 14
Centrifuge
Swinging bucket rotor is required that accepts flat-bottomed 250 mL bottles and is capable of spinning at up to 3,000 x g
Sanitization
Provided non-sterile; may be sanitized by filtering 70% ethanol through the device prior to use.
Jumbosep Device Reduces Processing
Time Over Other Devices

1 mg/mL BSA solution was processed in each of the above
devices until a 15-fold concentration was achieved.
Concentration Time

Concentrate dilute protein samples in less than 30 minutes
with 10, 30, and 100K Jumbosep devices.
Centrifuge
Swinging bucket rotor is required that accepts 250 mL bottles and is capable of spinning at up to 3000 x g
Sanitization
Provided non-sterile; may be sanitized by filtering 70% ethanol through the device prior to use
Testing info by rg