Materials of ConstructionFilter Media: Emflon II membrane (hydrophobic PVDF)
Housing: Polypropylene
Effective Filtration Area
20 cm2
Inlet/Outlet Connections
Stepped hose barbs, 6.4 - 12.7 mm (1/4 - 1/2 in.) diameter with slip luer ID in the hose barb
Maximum Operating Temperature
60 °C (140 °F)
Maximum Operating Pressure
4.1 bar (410 kPa, 60 psi) at ambient temperature
Recommended Integrity Test
Minimum Bubble Point -
60% IPA/40% Water (v:v)
0.8 bar (80 kPa, 12 psi)
Typical Air Flow Rate
27 L/min at 1.0 bar (100 kPa, 15 psi)
Endotoxin Level
< 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Bacterial Retention
Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838-83
Sterilization
Provided non-sterile. Stable with gamma irradiation up to 50 kilogray. Can be sterilized by autoclaving at 121 °C (250 °F) for 20 min.
Maximum autoclave stability, three cycles at 131 °C (268 °F) for 30 min.