Superior grade glass fiber filters for a variety of applications
- Type A/E meets the requirements for suspended solids testing, as described in Standard Methods for the Examination of Water and Wastewater, current edition.
- Reduce filtration costs and premature clogging when filtering difficult-to-filter or highly contaminated solutions.
- Extend filter life and make fewer final filter changes with high capacity prefilters.
- Eliminate sample contamination. Binder-free borosilicate glass fiber has no added extractables.
- Filter a wide range of particulate loads and viscous solutions with a selection of filter thicknesses from which to choose.
- Excellent wet strength for easy handling and filter integrity.
Used in a variety of sample cleanup, prefiltration, and analytical testing applications. Choose between binder-free borosilicate glass for complete purity or glass fiber with acrylic binder for added strength.
Type A/E Glass Fiber
- For testing dissolved and suspended solids in wastewater and gravimetric analysis of air pollutants.
- High flow rates, wet strength, and dirt (solids) holding capacities.
Type A/B Glass Fiber
- High dirt-loading capacity with 2.5 times thicker glass than Type A/C.
- Manufactured of the highest quality borosilicate glass microfibers.
Type A/C Glass Fiber
- For testing dissolved and suspended solids in wastewater.
- 254 µm (10 mils) thickness is useful for cell harvesting applications.
- Purity eliminates risk of unwanted contaminants leaching into the filtrate.
Type A/D Glass Fiber
- Excellent prefilters for solutions with a heavy load of large-sized particulate that must be removed.
- Large nominal pore size (3 µm) reduces membrane clogging.
Extra Thick Glass Fiber
- Preferred for prefiltration of viscous biological solutions.
- High particulate-holding capacity makes discs efficient depth filters and allows for filtration of large volumes of solutions.
Metrigard Glass Fiber
- Useful for prolonging membrane filter life in liquid systems that contain substantial amounts of particulate matter.
| Specifications and Selection Chart |
| Description |
Type A/E |
Type A/B |
Type A/C |
Type A/D |
Extra Thick |
Metrigard Discs |
| Typical Applications |
Water solids testing, air monitoring, gravimetric analysis |
Diagnostic applications, sample prefiltration |
Cell harvesting, prefiltration, solids testing |
Prefiltration of solutions with large-sized particulate |
Prefiltration of heavily contaminated samples |
Prefiltration in systems with high particulate matter |
| Filter Media |
Borosilicate glass without binder |
Borosilicate glass without binder |
Borosilicate glass without binder |
Borosilicate glass without binder |
Glass fiber with acrylic binder** |
Ultrafine glass fiber with acrylic binder** |
| Pore Size (nominal) |
1 µm |
1 µm |
1 µm |
3 µm |
1 µm |
0.5 µm |
| Typical Thickness |
330 µm (13 mils) |
660 µm (26 mils) |
254 µm (10 mils) |
660 µm (26 mils) |
1270 µm (50 mils) |
330 µm (13 mils) |
| Typical Water Flow Rate mL/min/cm2 at 0.3 bar (30 kPa, 5 psi) |
250 |
124 |
153 |
649 |
210 |
80 |
| Typical Air Flow Rate L/min/cm2 at 0.7 bar (70 kPa, 10 psi) |
68 |
24 |
40 |
139 |
26 |
21 |
| Maximum Operating Temperature |
Air 550° C (1022° F) |
Air 550° C (1022° F) |
Air 550° C (1022° F) |
Air 550° C (1022° F) |
Water 135° C (275° F) |
Water 135° C (275° F) |
| Sterilization |
Autoclavable |
Autoclavable |
Autoclavable |
Autoclavable |
Autoclavable |
Autoclavable |
| Typical Aerosol Retention* |
99.98% |
- |
- |
- |
99.97% |
99.99% |
*Following ASTM D 2986-95A 0.3 µm (DOP) at 32 L/min/100 cm2 filter media. **Binder is 5% of total material. |
Contact Customer Service at 800-521-1520 for possible lead-time and date item will be available to ship.
|
 |